The Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes: Comparison of an automated oscillometric ABI and standard hand-held doppler ABI

Khan, Shair Zaman and Haque, Muhammad Saiful and Zafar, Awn Bin and Waris, Nazish and Miyan, Zahid and Haque, Muhammad Saiful and Fawwad, Asher (2019) The Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes: Comparison of an automated oscillometric ABI and standard hand-held doppler ABI. Pakistan Journal of Medical Sciences, 35 (4). ISSN 1682-024X

[thumbnail of 30-Article Text-6762-1-10-20190710.pdf] Text
30-Article Text-6762-1-10-20190710.pdf - Published Version

Download (453kB)

Abstract

The Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes Comparison of an automated oscillometric ABI and standard hand-held doppler ABI Shair Zaman Khan Muhammad Saiful Haque Awn Bin Zafar Nazish Waris Zahid Miyan Muhammad Saiful Haque Asher Fawwad

Objective: To compare the difference between an automated oscillometric ABI measurement as compared to standard hand-held doppler ABI in patients with Type-2 diabetes. Methods: This prospective study was conducted at foot clinic of Baqai Institute of Diabetology and Endocrinology (BIDE), Baqai Medical University (BMU), a tertiary care unit, Karachi-Pakistan. The duration of study was February 2018 to March 2018. Patients with Type-2 diabetes attending the outpatient department (OPD) of foot clinic, irrespective of their symptoms were included. Baseline demographic, anthropometric measurements and biochemical parameters were recorded. The ABI was calculated with both devices by an automated oscillometric machine and standard hand-held doppler with the same investigator. Results: Total of 93 patients with Type-2 diabetes, 18 (19.4%) females and 75(80.6%) males were recruited. Mean age was 54.67±9.59 years and mean systolic/diastolic blood pressure was 131.38±20.2/ 80.36±10.23mmHg. Most of the patients had poor glycemic control at presentation with a mean HbA1c of 9.56±2.44%. Mean standard handheld doppler ABI and automated oscillometric ABI was 1.28±1.08 and 1.07±0.23 for right foot (mean difference = 0.21; P= 0.075), and 1.14±0.45 and 1.1±0.25 for left foot (mean difference =0.04; P=0.434), respectively. Similarly, sensitivity and specificity between two modalities was observed 60% and 93.90% for right foot, meanwhile, 60% and 97.40% for left foot, respectively. Conclusion: An automated oscillometric method is comparable with standard handheld-doppler method. It is cost effective, convenient and less time consuming, can be widely used to measure ABI without special training. doi: https://doi.org/10.12669/pjms.35.4.30 How to cite this:Khan SZ, Awn-Bin-Zafar, Waris N, Miyan Z, Ulhaque MS, Fawwad A. Comparison of ankle-brachial index (ABI) measured by an automated oscillometric apparatus with that by standard hand-held doppler in patients with Type-2 diabetes. Pak J Med Sci. 2019;35(5):1167-1172. doi: https://doi.org/10.12669/pjms.35.4.30 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
07 10 2019 10.12669/pjms.35.4.30 http://www.pjms.org.pk/index.php/pjms/article/view/30 http://www.pjms.org.pk/index.php/pjms/article/download/30/231 http://www.pjms.org.pk/index.php/pjms/article/download/30/231

Item Type: Article
Subjects: STM Academic > Medical Science
Depositing User: Unnamed user with email support@stmacademic.com
Date Deposited: 25 Apr 2023 08:18
Last Modified: 23 Mar 2024 04:33
URI: http://article.researchpromo.com/id/eprint/610

Actions (login required)

View Item
View Item